Changes between Initial Version and Version 1 of Ticket #627, comment 1
- Timestamp:
- Jun 5, 2018, 3:06:57 PM (6 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
Ticket #627, comment 1
initial v1 3 3 -The fenix sensor needed a replacement of the array so the new wavelengths are expected to be different 4 4 5 -Specim performed its own calibration on Feb 2018. On a first look the new hdr files by Specim looks to be in accordance with our wavelength calibration 5 -Specim performed its own calibration on Feb 2018. On a first look the new hdr files by Specim looks to be in accordance with our wavelength calibration (close to the scale given by Fenix) 6 6 7 7 -Details recorded are within the spreadsheet: ./data_details/recorded_at_BAS_lab/Capture_Spreadsheet_2018-04.ods … … 10 10 Day1: Fenix wavelength cal+radiometric cal (no dark frames for 1 spectral lamp and some radiometric files). Also some Owl validation data with the black body. 11 11 Day2: Monochromator tests for different ranges of wavelengths in the SWIR +1 in the VNIR (no Fenix calibration data) 12 Day3: Fenix wavelength cal+radiometric cal .12 Day3: Fenix wavelength cal+radiometric cal (no black cloth used). 13 13 Day4: Fenix wavelength cal+radiometric cal. Also some Owl validation data with the black body.