Changes between Initial Version and Version 1 of Ticket #627, comment 1


Ignore:
Timestamp:
Jun 5, 2018, 3:06:57 PM (6 years ago)
Author:
asm
Comment:

Legend:

Unmodified
Added
Removed
Modified
  • Ticket #627, comment 1

    initial v1  
    33-The fenix sensor needed a replacement of the array so the new wavelengths are expected to be different
    44
    5 -Specim performed its own calibration on Feb 2018. On a first look the new hdr files by Specim looks to be in accordance with our wavelength calibration
     5-Specim performed its own calibration on Feb 2018. On a first look the new hdr files by Specim looks to be in accordance with our wavelength calibration (close to the scale given by Fenix)
    66
    77-Details recorded are within the spreadsheet: ./data_details/recorded_at_BAS_lab/Capture_Spreadsheet_2018-04.ods
     
    1010    Day1: Fenix wavelength cal+radiometric cal (no dark frames for 1 spectral lamp and some radiometric files). Also some Owl validation data with the black body.
    1111    Day2: Monochromator tests for different ranges of wavelengths in the SWIR +1 in the VNIR (no Fenix calibration data)
    12     Day3: Fenix wavelength cal+radiometric cal.
     12    Day3: Fenix wavelength cal+radiometric cal (no black cloth used).
    1313    Day4: Fenix wavelength cal+radiometric cal. Also some Owl validation data with the black body.