    v18 v19  
    3232 * fast_bil_median=<path to fast_bil_median> (%(scriptdir)s/fast_bil_median)
    3333 * fityk=<path to cfityk>(cfityk)
    34  * linefile=<path to CSV file containing spectral lines, the wavelengths of each spectral peak that each line provide> (Can be edited for new lamps)
     34 * linefile=<path to CSV file containing spectral lines, the wavelengths of each spectral peak that each line provide> (Can be edited for new lamps. I advice to use the Mylar filter only for checking the spectra in the SWIR is correct comparing with an scale not using the Mylar filter. The Mylar filter has an accuracy of 1nm and introduces a bigger than real FWHM so do not use it for the final version)
    3535 * lamp_index_file=<path to CSV file matching data file names to spectral lamps, usually named as lamp_lookup.csv>
    3636A example of the lamp_index file is: