Changes between Version 18 and Version 19 of Procedures/OwlProcessing


Ignore:
Timestamp:
Dec 7, 2015, 6:16:44 PM (9 years ago)
Author:
lah
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • Procedures/OwlProcessing

    v18 v19  
    1010
    1111The radiometric calibration will fail if the dark frames are saved in a separate file to the data (Ops sometimes do this). Check the capture directory for any extra files labelled like they might be dark frames and append them to the end of the data file. Remember to add the autodarkstartline key to the header file. Alternatively use the script stitchOwl.py.
     12
     13Specim's calibration tool actually only uses dark frames to detect blinkers; it does not use them to radiometrically calibrate the data. Therefore, if dark frames are completely missing a calibration file (e.g. T1) may be used instead.
    1214
    1315== Radiometric Calibration ==