Changes between Version 18 and Version 19 of Procedures/OwlProcessing
- Timestamp:
- Dec 7, 2015, 6:16:44 PM (9 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
Procedures/OwlProcessing
v18 v19 10 10 11 11 The radiometric calibration will fail if the dark frames are saved in a separate file to the data (Ops sometimes do this). Check the capture directory for any extra files labelled like they might be dark frames and append them to the end of the data file. Remember to add the autodarkstartline key to the header file. Alternatively use the script stitchOwl.py. 12 13 Specim's calibration tool actually only uses dark frames to detect blinkers; it does not use them to radiometrically calibrate the data. Therefore, if dark frames are completely missing a calibration file (e.g. T1) may be used instead. 12 14 13 15 == Radiometric Calibration ==